SourceMed · Shelf #9591 in the ask interface
Pulse CO-oximetry for carboxyhaemoglobin and methaemoglobin vs laboratory CO-oximetry (4 studies, documented N = 145 patients/volunteers plus 120 ED patients in the clinical study): in the emergency department the Rad-57 missed half of patients above 15% carboxyhaemoglobin (sensitivity 48%, labelled in words) with limits of agreement −11.6 to +14.4 points, and volunteer accuracy held only while oxygen saturation stayed above about 85%
Shelf #9591 · Critical care & sepsis
MEASUREMENT ACCURACY — NOT ESTABLISHED for interchangeable use with laboratory CO-oximetry: in the only emergency-department study the device detected only 48% of patients with laboratory carboxyhaemoglobin above 15% (sensitivity, labelled in words at first occurrence; specificity 99%), with Bland–Altman limits of agreement of −11.6 to +14.4 percentage points against a prespecified acceptable ±5; controlled volunteer studies show small average bias for SpCO and for the revised SpMet algorithm, but the original SpMet readings overestimated by up to 10–40 points as saturation fell below 95%, and SpCO is not reported at all below about 85% saturation
Stamped: 2026-11-28.
Evidence published: 2010–2013.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
How accurately do multi-wavelength pulse CO-oximeters (Masimo Rainbow / Rad-57 and Radical-7 — named as published) measure carboxyhaemoglobin (SpCO) and methaemoglobin (SpMet) against laboratory blood CO-oximetry — in volunteers under controlled conditions and in emergency-department patients with suspected carbon monoxide poisoning?
Balance of the evidence
MEASUREMENT ACCURACY — NOT ESTABLISHED for interchangeable use — Composition: 4 primary studies (1 emergency-department cross-section of 120 patients; 3 controlled volunteer studies, n = 12, 12 and 14). Touger 2010 (120 ED patients with suspected carbon monoxide poisoning, Rad-57 against simultaneous laboratory carboxyhaemoglobin; median laboratory value 2.3%, range 0–38%): mean difference 1.4 percentage points, but limits of agreement were −11.6 to +14.4 points — beyond the ±5 the investigators prespecified as clinically acceptable — and the device identified 11 of 23 patients with laboratory values above 15% (sensitivity 48%, 95% CI 27–69%; specificity 99%). Feiner 2013 (12 volunteers, induced carboxyhaemoglobinaemia to 12% with graded hypoxaemia): SpCO bias was −0.98 ± 2.6 points at saturation 95% or above and −0.7 ± 4.0 below it, but below about 85% saturation the device returned only low-signal errors and no SpCO value. Feiner 2010 (methaemoglobin, original algorithm; 14 volunteers): SpMet bias and precision worsened steeply as saturation fell (24.8 ± 15.6 points of bias at saturation 70–80%). Feiner & Bickler 2010 (revised algorithm; 12 volunteers): SpMet bias 0.16% with precision 0.83% across saturations 74–100% and methaemoglobin 0–14%. MEASUREMENT ACCURACY — NOT ESTABLISHED for interchangeable use: under laboratory conditions with the current algorithm the average error is small, but the emergency-department agreement limits are several times wider than clinically acceptable and half of significantly poisoned patients were missed at the 15% threshold — a screening reading, not a measurement. Sensitivity and specificity are labelled in words at first occurrence in this balance, per the labelling rule.
MeSH: Oximetry · Carboxyhemoglobin · Methemoglobin · Carbon Monoxide Poisoning · Sensitivity and Specificity
Cards (5)
Card 1 — In 120 ED patients the Rad-57 found only 11 of 23 with laboratory carboxyhaemoglobin above 15% (sensitivity 48%; specificity 99%), with limits of agreement −11.6 to +14.4 points (Touger 2010)
Study: Primary accuracy study (device vs laboratory CO-oximetry in ED patients; n = 120)
Population: A cross-sectional cohort study of a convenience sample of 120 adult and paediatric emergency-department patients with suspected carbon monoxide poisoning at a Level I trauma, burn and hyperbaric centre; Rad-57 pulse CO-oximeter carboxyhaemoglobin was measured simultaneously with arterial or venous laboratory carboxyhaemoglobin (median 2.3%, interquartile range 1–8.5%, range 0–38%). Sensitivity and specificity are labelled in words here at first occurrence on this shelf, per the shelf's labelling rule.
Summary: In 120 ED patients the Rad-57 found only 11 of 23 with laboratory carboxyhaemoglobin above 15% (sensitivity 48%; specificity 99%), with limits of agreement −11.6 to +14.4 points (Touger 2010)
Result: The mean difference (laboratory minus device) was 1.4 percentage points (95% CI 0.2 to 2.6), but Bland–Altman limits of agreement were −11.6% and +14.4% carboxyhaemoglobin — exceeding the ±5% defined a priori as clinically acceptable. The device correctly identified 11 of 23 patients with laboratory values above 15% (sensitivity 48%, 95% CI 27% to 69%) and gave one false positive among 97 patients at or below 15% (specificity 99%, 95% CI 94% to 100%). The authors concluded the measurements may not be used interchangeably.
Watch out: A convenience sample at one referral centre with most values low (median 2.3%) — the wide limits of agreement are estimated across the full 0–38% range; the 48% sensitivity rests on 23 patients above the threshold, with the wide confidence interval printed.
Source: Touger M, Birnbaum A, Wang J et al. Performance of the RAD-57 pulse CO-oximeter compared with standard laboratory carboxyhemoglobin measurement. Annals of emergency medicine. 2010. PMID: 20605259. Registry ID: not stated in the record. MeSH: Adolescent · Adult · Carbon Monoxide Poisoning · Carboxyhemoglobin · Cross-Sectional Studies · Emergency Service, Hospital · Female · Humans · Male · Middle Aged · Oximetry · Point-of-Care Systems · Sensitivity and Specificity · Young Adult
Card 2 — In 12 volunteers SpCO bias stayed near −1 point down to 85% saturation — below which the device reported nothing at all (Feiner 2013, N = 12 printed on the face)
Study: Primary accuracy study (controlled volunteer induction study; n = 12 — N printed on the face of this card)
Population: Twelve healthy non-smoking adult volunteers wore two standard pulse-oximeter probes and two Rainbow CO-oximetry probes with a radial arterial catheter, through three interventions: graded hypoxaemia (saturation 100% down to 80%), induced carboxyhaemoglobinaemia to 12% at normoxia, and the two combined (about 25 samples per subject).
Summary: In 12 volunteers SpCO bias stayed near −1 point down to 85% saturation — below which the device reported nothing at all (Feiner 2013, N = 12 printed on the face)
Result: The device detected hypoxaemia accurately with normal and raised carboxyhaemoglobin (bias 0.44 ± 1.69 points at carboxyhaemoglobin under 4%; −0.29 ± 1.64 at 4% or above). Carboxyhaemoglobin itself was measured with bias −0.98 ± 2.6 points at saturation 95% or above and −0.7 ± 4.0 below 95% (root-mean-square error 2.8% vs 4.0%) — but when saturation fell below approximately 85%, the device always gave low-signal-quality errors and reported no SpCO value.
Watch out: Twelve healthy volunteers — N printed on the face — under laboratory conditions with induced, stable levels; this is the device's best case, and it still falls silent exactly when saturation drops below about 85%, a range real poisoned and hypoxaemic patients occupy.
Source: Feiner JR, Rollins MD, Sall JW et al. Accuracy of carboxyhemoglobin detection by pulse CO-oximetry during hypoxemia. Anesthesia and analgesia. 2013. PMID: 23477959. Registry ID: not stated in the record. MeSH: Adolescent · Adult · Carbon Monoxide Poisoning · Carboxyhemoglobin · Female · Humans · Hypoxia · Male · Middle Aged · Oximetry · Young Adult
Card 3 — Original SpMet algorithm: bias grew from 1.9 points at saturation 95–100% to 24.8 points at 70–80%, with readings overestimating by 10–40 points (Feiner 2010, N = 14 printed on the face)
Study: Primary accuracy study (controlled volunteer induction study, original algorithm; n = 14 — N printed on the face of this card)
Population: Two groups of healthy adults (8 and 6) wore multiple sensors with a radial arterial catheter while intravenous sodium nitrite raised methaemoglobin toward 7–8% (first group) or 15% (second group) and inspired oxygen was varied to give saturations from 66% to 100%; device readings were compared with a Radiometer multi-wavelength laboratory oximeter over 170 blood draws.
Summary: Original SpMet algorithm: bias grew from 1.9 points at saturation 95–100% to 24.8 points at 70–80%, with readings overestimating by 10–40 points (Feiner 2010, N = 14 printed on the face)
Result: Across the full saturation span, SpMet bias and precision were 7.7 ± 13.0 points. Accuracy was best at saturation 95–100% (bias 1.9 ± 2.5 points) and worst at 70–80% (24.8 ± 15.6 points), with readings in error by more than 5 points becoming more frequent with each 5-point fall in saturation (P < 0.05); at low saturations the device overestimated true methaemoglobin by 10 to 40 points at times. Elevated methaemoglobin also biased the device's oxygen-saturation reading downward (−6.3 ± 3.0 points at saturation 95–100% with 4–8.3% methaemoglobin), similar to a conventional two-wavelength oximeter.
Watch out: Fourteen volunteers — N printed on the face — and the original manufacturer algorithm; carded because its failure mode (gross overestimation as saturation falls) is what the manufacturer's revision then had to fix, which the next card measures.
Source: Feiner JR, Bickler PE, Mannheimer PD. Accuracy of methemoglobin detection by pulse CO-oximetry during hypoxia. Anesthesia and analgesia. 2010. PMID: 20007731. Registry ID: not stated in the record. MeSH: Adult · Blood Gas Analysis · Carbon Monoxide · Carboxyhemoglobin · Data Interpretation, Statistical · Female · Hemoglobinometry · Humans · Hypoxia · Male · Methemoglobin · Methemoglobinemia · Oximetry · Oxygen · Predictive Value of Tests · Reproducibility of Results · Sodium Nitrite
Card 4 — After the manufacturer's modification, SpMet bias was 0.16 points with precision 0.83 across saturations 74–100% and methaemoglobin 0–14% (Feiner & Bickler 2010, N = 12 printed on the face)
Study: Primary accuracy study (controlled volunteer induction study, revised algorithm; n = 12 — N printed on the face of this card)
Population: Twelve healthy adult volunteers wore sensors on each hand with a radial arterial catheter while intravenous sodium nitrite (about 300 mg) raised methaemoglobin to a 7–11% target and inspired oxygen varied saturation from 70% to 100%; 307 blood draws gave 602 paired readings against a Radiometer ABL800 FLEX laboratory oximeter.
Summary: After the manufacturer's modification, SpMet bias was 0.16 points with precision 0.83 across saturations 74–100% and methaemoglobin 0–14% (Feiner & Bickler 2010, N = 12 printed on the face)
Result: With the modified device, methaemoglobin bias and precision over the full saturation span were 0.16% and 0.83%, similar across the span — acceptably accurate over saturations 74–100% and methaemoglobin 0–14.4%, correcting the spurious high readings during hypoxia reported for the earlier version. The device's oxygen-saturation readings remained biased −1.93 points across the 70–100% range in the presence of elevated methaemoglobin.
Watch out: Twelve volunteers — N printed on the face — from the same laboratory that documented the original failure; an independent clinical-population validation of the revised SpMet algorithm is not among the carded records, and the emergency-department SpCO study above used the clinical device as deployed.
Source: Feiner JR, Bickler PE. Improved accuracy of methemoglobin detection by pulse CO-oximetry during hypoxia. Anesthesia and analgesia. 2010. PMID: 20841412. Registry ID: not stated in the record. MeSH: Administration, Inhalation · Biomarkers · Carboxyhemoglobin · Female · Fingers · Hemoglobins · Humans · Hypoxia · Injections, Intravenous · Male · Materials Testing · Methemoglobin · Oximetry · Oxygen · Predictive Value of Tests · Reproducibility of Results · San Francisco · Sodium Nitrite
Card 5 — Synthesis: small average bias in the laboratory, unacceptably wide agreement in the department — SpCO/SpMet readings screen, they do not measure
Study: Synthesis of the carded studies (4 primary studies; 120 patients + 38 volunteers)
Population: Synthesis of the four carded primary studies (120 ED patients; 38 volunteers across three induction studies, faces printed).
Summary: Synthesis: small average bias in the laboratory, unacceptably wide agreement in the department — SpCO/SpMet readings screen, they do not measure
Result: Controlled studies show the technology can track induced carboxyhaemoglobin and — after revision — methaemoglobin with small average bias, within limits: no SpCO output below about 85% saturation, and a documented earlier failure mode of 10–40-point SpMet overestimates in hypoxia. The one real-world study found agreement limits (−11.6 to +14.4 points) several times the prespecified acceptable width and sensitivity of 48% at the 15% carboxyhaemoglobin threshold. A reassuring SpCO cannot exclude significant poisoning; a blood measurement remains the deciding test.
Watch out: This shelf covers the dyshaemoglobin channels (SpCO, SpMet) of pulse CO-oximetry only. The same platform's haemoglobin channel (SpHb) is a different measurement with its own live shelf (#8816) and its own agreement data, and nothing here should be read onto it.
Source: Synthesis of the carded records above (PMIDs 20605259, 23477959, 20007731, 20841412). Registry ID: not stated in the records. MeSH: Oximetry · Carboxyhemoglobin · Methemoglobin · Carbon Monoxide Poisoning · Sensitivity and Specificity