SourceMed · Shelf #9645 in the ask interface
TricValve bicaval caval valve implantation for severe symptomatic tricuspid regurgitation (TRICUS programme, single-arm; documented unique N = 44): 95.5% met a clinical-improvement composite at 1 year, with 3 deaths and a 29.5% heart-failure rehospitalisation rate — no randomised comparison exists
Shelf #9645 · Heart, vessels, blood pressure & clotting
POSSIBLE (benefit) — a single-arm programme of 44 unique patients with consistent quality-of-life and functional-class gains and abolished hepatic-vein backflow in about two-thirds, but no control group of any kind, so improvement over medical therapy alone is unproven
Stamped: 2026-11-28.
Evidence published: 2022–2024.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In patients with severe symptomatic tricuspid regurgitation and significant caval backflow who are ineligible for open heart surgery, does heterotopic bicaval valve implantation with the TricValve system improve quality of life, functional class and congestion compared with the patients' own baseline?
Balance of the evidence
POSSIBLE (benefit) — Composition: 1 single-arm device programme reported twice (a 35-patient European cohort at 6 months and the combined 44-patient cohort at 1 year; the 35 are contained within the 44, so the documented unique N is 44, not 79). At 1 year, 42 of 44 patients (95.5%) met the clinical-improvement composite (Kansas City Cardiomyopathy Questionnaire gain ≥15 points, NYHA class I–II, or 6-minute walk gain ≥40 m); 3 patients (6.8%) had died and the heart-failure rehospitalisation rate was 29.5%. Hepatic-vein backflow was abolished and persisted in 63.8%. Boundary: live #715 (edge-to-edge tricuspid repair, TRILUMINATE) and live #2250 (orthotopic tricuspid valve replacement, TRISCEND II) treat the valve itself in patients suitable for those approaches; this shelf is the heterotopic caval option for patients ineligible for them, a different intervention and question. The tested product is the TricValve Transcatheter Bicaval Valves System (Products + Features), named as published. POSSIBLE (benefit), single-arm evidence only.
MeSH: Aged · Aged, 80 and over · Cardiac Catheterization · Cardiomyopathies · Female · Heart Valve Prosthesis Implantation · Humans · Male · Prospective Studies · Quality of Life · Severity of Illness Index · Treatment Outcome · Tricuspid Valve · Tricuspid Valve Insufficiency · Follow-Up Studies
Cards (3)
Card 1 — TRICUS EURO at 6 months: quality of life and functional class improved after TricValve implantation (Estévez-Loureiro 2022, n = 35, single-arm)
Study: Prospective single-arm multicentre study (n = 35 treated)
Population: 35 patients (mean age 76 ± 6.8 years, 83% women) with severe symptomatic tricuspid regurgitation despite optimal medical therapy, at high surgical risk, treated with the TricValve system (Products + Features, named as published) at 12 European centres; all were in NYHA functional class III or IV at baseline. Prospective, nonblinded, nonrandomised, single-arm trial; the primary endpoint was quality-of-life improvement (Kansas City Cardiomyopathy Questionnaire and NYHA class) at 6 months.
Summary: TRICUS EURO at 6 months: quality of life and functional class improved after TricValve implantation (Estévez-Loureiro 2022, n = 35, single-arm)
Result: Procedural success was 94% at 30 days, with no procedural deaths or conversions to surgery. The Kansas City Cardiomyopathy Questionnaire score rose from 42.01 ± 22.3 at baseline to 59.7 ± 23.6 at 6 months (P = 0.004), and 79.4% of patients were in NYHA class I or II at 6 months (P = 0.0006). Six-month all-cause mortality was 8.5% and heart-failure hospitalisation 20%.
Watch out: No comparator: symptom and quality-of-life gains after an invasive procedure in an unblinded single-arm study are the outcomes most exposed to placebo and expectation effects, and 8.5% of the cohort had died by 6 months. This cohort is contained within the 44-patient 1-year report carded next — the two reports are one programme, not two studies.
Source: Estévez-Loureiro R et al. 6-Month Outcomes of the TricValve System in Patients With Tricuspid Regurgitation: The TRICUS EURO Study. JACC. Cardiovascular Interventions. 2022. PMID: 35583363. Registry ID: not stated in the record. MeSH: Aged · Aged, 80 and over · Cardiac Catheterization · Cardiomyopathies · Female · Heart Valve Prosthesis Implantation · Humans · Male · Prospective Studies · Quality of Life · Severity of Illness Index · Treatment Outcome · Tricuspid Valve · Tricuspid Valve Insufficiency
Card 2 — TRICUS programme at 1 year: 95.5% met the clinical-improvement composite (Blasco-Turrión 2024, combined n = 44, single-arm)
Study: Prospective single-arm programme report, combined cohorts (n = 44 treated)
Population: 44 patients (mean age 76.2 ± 7.5 years, 81% women; TRISCORE 5.3 ± 1.3) with NYHA class III–IV severe tricuspid regurgitation, optimally medicated, ineligible for open heart surgery and with significant caval backflow, from the combined TRICUS and TRICUS EURO single-arm trials of the TricValve system (Products + Features). The primary endpoint was quality-of-life metrics and functional status at 1 year.
Summary: TRICUS programme at 1 year: 95.5% met the clinical-improvement composite (Blasco-Turrión 2024, combined n = 44, single-arm)
Result: Clinical improvement at 1 year was achieved in 42 patients (95.5%), defined as at least one of: Kansas City Cardiomyopathy Questionnaire gain ≥15 points, improvement to NYHA class I or II, or 6-minute walk gain ≥40 m. Three patients (6.8%) died by 1 year (1 cardiovascular) and the heart-failure rehospitalisation rate was 29.5%. Hepatic-vein backflow was abolished and persisted in 63.8% of patients, alongside a reduction in NT-proBNP levels (P = 0.032) — a measured marker change, stated neutrally — and reduced diuretic treatment. No stent fracture, conduction disturbance or clinically significant leaflet thrombosis was detected.
Watch out: The composite counts a patient as improved on any one of three soft measures, and there is still no control arm: how much of the gain exceeds what optimised medical care and follow-up alone would produce is unknown. Randomised trials of caval implantation against medical therapy are described as ongoing in the literature, not reported.
Source: Blasco-Turrión S et al. Bicaval TricValve Implantation in Patients With Severe Symptomatic Tricuspid Regurgitation: 1-Year Follow-Up Outcomes. JACC. Cardiovascular Interventions. 2024. PMID: 38069986. Registry ID: ClinicalTrials.gov NCT03723239. MeSH: Aged · Aged, 80 and over · Female · Humans · Male · Cardiac Catheterization · Follow-Up Studies · Heart Valve Prosthesis Implantation · Prospective Studies · Quality of Life · Severity of Illness Index · Treatment Outcome · Tricuspid Valve · Tricuspid Valve Insufficiency
Card 3 — Synthesis: consistent uncontrolled gains in a no-option population — POSSIBLE (benefit)
Study: Synthesis
Population: Documented unique N = 44 (the 35-patient European 6-month report is a subset of the combined 44-patient 1-year report; the Ns are not added).
Summary: Synthesis: consistent uncontrolled gains in a no-option population — POSSIBLE (benefit)
Result: Across both reports the direction is consistent: large quality-of-life gains, most patients reaching NYHA class I–II, abolished hepatic backflow in about two-thirds, against 6.8% 1-year mortality and 29.5% heart-failure rehospitalisation in a cohort with no surgical option. Every efficacy number is a within-patient change without a comparator.
Watch out: Single-arm shelves cannot separate device effect from selection, co-care and expectation. The stamp is POSSIBLE and the tier is C for exactly that reason; a randomised result would supersede this shelf's basis.
Source: Synthesis of the carded records above (PMIDs 35583363, 38069986).