SourceMed · Shelf #9689 in the ask interface

Home spirometry versus clinic spirometry (3 studies, documented N = 2,975): unsupervised home FEV1 did not agree with clinic values in 2,857 trial patients, while virtually supervised home testing reached intraclass correlation 0.95 in 68 patients

Shelf #9689 · Lungs, breathing & sleep

MEASUREMENT ACCURACY — NOT ESTABLISHED for unsupervised interchangeable use with clinic spirometry; supervised home testing showed high agreement in one small study

Stamped: 2026-11-28.

Evidence published: 2007–2025.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

How closely do home spirometry measurements (FEV1, FVC, peak flow) agree with supervised clinic or laboratory spirometry in asthma and COPD — and does supervision, human or AI over-reading, close the gap enough for the two to be used interchangeably?

Balance of the evidence

MEASUREMENT ACCURACY — NOT ESTABLISHED for unsupervised interchangeable use with clinic spirometry — Composition: 3 primary agreement studies (documented N = 2,975: 2,857 + 68 + 50). No carded record prints a sensitivity/specificity pair (this is agreement, not diagnosis), and none is invented; agreement is reported as Bland-Altman findings and intraclass correlations as the records give them. The largest evidence is discouraging for unsupervised use: in post hoc analysis of two phase 3/2B trials (2,436 + 421 patients with uncontrolled asthma), home trough FEV1 improvements were smaller and less consistent than clinic measurements, and Bland-Altman plots showed poor agreement between home and clinic FEV1 at baseline and week 24 — the authors conclude unsupervised home readings are not interchangeable with clinic measurements (Oppenheimer 2023). Supervision changes the picture: in the COMPAIR study (68 adults with asthma or COPD), virtually supervised home spirometry on a named portable device agreed with directly supervised clinic spirometry at intraclass correlation 0.95 for both FEV1 and FVC, rising to 0.98 and 0.96 with AI over-reading (Pradhan 2025). For scale: an intraclass correlation of 0.70 suffices for comparing groups, while 0.90 is the ideal — a measure is acceptable for individual-patient decisions when it reaches the 90% threshold — so 0.95-0.98 clears the individual-decision bar in that small supervised study, while the unsupervised trial data provide no such figure and failed agreement testing outright. In 50 children, a home spirometer read peak flow lower than the laboratory pneumotachograph by a mean difference whose 95% CI was 14-30 L/min, with individual differences over 100 L/min, and FEV1 lower by 0.02-0.1 L (Brouwer 2007). MEASUREMENT ACCURACY — NOT ESTABLISHED for unsupervised interchangeable use: the default home reading cannot be swapped for a clinic value; agreement at the individual-decision standard has been shown only for virtually supervised testing with quality over-reading, in one 68-patient study.

MeSH: Humans · Administration, Inhalation · Asthma · Bronchodilator Agents · Chlorobenzenes · Double-Blind Method · Fluticasone · Forced Expiratory Volume · Lung · Nebulizers and Vaporizers · Spirometry · Treatment Outcome · Benzyl Alcohols · Adolescent · Child · Female · Male · Peak Expiratory Flow Rate · Reproducibility of Results

Cards (4)

Card 1 — 2,857 asthma trial patients: home trough FEV1 showed poor Bland-Altman agreement with clinic FEV1 and smaller, less consistent changes (Oppenheimer 2023)

Study: Primary agreement study (home vs clinic spirometry, post hoc trial analysis; n = 2,857)

Population: Post hoc analysis of two randomised trials in uncontrolled asthma: CAPTAIN (2,436 patients) and study 205832 (421 patients); trough FEV1 measured by home spirometry and by supervised in-person clinic spirometry.

Summary: 2,857 asthma trial patients: home trough FEV1 showed poor Bland-Altman agreement with clinic FEV1 and smaller, less consistent changes (Oppenheimer 2023)

Result: Treatment-related FEV1 improvements were seen with both methods, but improvements measured at home were of lower magnitude and less consistent than clinic measurements. Bland-Altman plots suggested poor agreement between home and clinic trough FEV1 at baseline and at week 24. The authors conclude home spirometry was less consistent than, and lacked agreement with, clinic spirometry, so unsupervised home readings are not interchangeable with clinic measurements — with the caveat that the finding may apply only to the specific device and coaching methods used in these trials.

Watch out: A post hoc analysis, and the record prints the poor-agreement conclusion without numeric limits of agreement — no limits are reconstructed here; device and coaching were those of the parent trials.

Source: Oppenheimer J, Hanania NA, Chaudhuri R et al.. Clinic vs Home Spirometry for Monitoring Lung Function in Patients With Asthma. Chest. 2023. PMID: 37385337. Registry ID: ClinicalTrials.gov NCT02924688 and NCT03012061 (the source trials, as printed in the record). MeSH: Humans · Administration, Inhalation · Asthma · Bronchodilator Agents · Chlorobenzenes · Double-Blind Method · Fluticasone · Forced Expiratory Volume · Lung · Nebulizers and Vaporizers · Spirometry · Treatment Outcome · Benzyl Alcohols

Card 2 — Virtually supervised home spirometry: intraclass correlation 0.95 for FEV1 and FVC against clinic testing, 0.98 with AI over-read (Pradhan 2025; n = 68)

Study: Primary agreement study (supervised home vs clinic spirometry; n = 68)

Population: 68 adults with physician-diagnosed asthma (47) or COPD (21) in UK primary/secondary care; spirometry on the NuvoAir Air Next portable spirometer (named as published), directly supervised in clinic on days 1 and 8 and virtually supervised at home on days 2 and 7, with manual and ArtiQ.QC AI over-reading.

Summary: Virtually supervised home spirometry: intraclass correlation 0.95 for FEV1 and FVC against clinic testing, 0.98 with AI over-read (Pradhan 2025; n = 68)

Result: Intraclass correlation between clinic and home FEV1 was 0.95 (95% CI 0.92-0.97) and for FVC 0.95 (95% CI 0.92-0.97) with human over-reading, rising slightly with AI over-reading (FEV1 0.98, 95% CI 0.97-0.99; FVC 0.96, 95% CI 0.93-0.97); Bland-Altman plots showed good agreement, better with AI over-read. For scale, an intraclass correlation of 0.70 suffices for comparing groups and 0.90 is the ideal for individual-patient decisions, so these values clear the individual-decision threshold. Over 95% of patients found the device and app easy to use. This record carries no MeSH list, and none is reconstructed.

Watch out: One small 8-day study of one named device with healthcare-professional supervision and over-reading built in — it validates a supervised service model, not unsupervised home testing; the record carries no MeSH list.

Source: Pradhan A, Pollington J, Saralaya D et al.. COMPAIR: a comparison of directly supervised clinic versus virtually supervised home spirometry. ERJ open research. 2025. PMID: 41158491. Registry ID: not stated in the record. MeSH: none in the record

Card 3 — Children, n = 50: home spirometer peak flow read 14-30 L/min lower than the laboratory pneumotachograph, with individual gaps over 100 L/min (Brouwer 2007)

Study: Primary agreement study (home spirometer vs laboratory pneumotachograph in children; n = 50)

Population: 50 stable asthmatic children aged 6-17 years (33 boys), performing peak expiratory flow and FEV1 on an electronic home spirometer (Koko Peak Pro, named as published) and on the hospital laboratory pneumotachograph in random order, all measurements meeting standard quality criteria under professional supervision.

Summary: Children, n = 50: home spirometer peak flow read 14-30 L/min lower than the laboratory pneumotachograph, with individual gaps over 100 L/min (Brouwer 2007)

Result: Peak flow on the home spirometer was considerably lower than on the pneumotachograph (95% CI for the difference 14-30 L/min, P < 0.0001), and individual differences could exceed 100 L/min. FEV1 was slightly but significantly lower at home (95% CI for the difference 0.02-0.1 L, P = 0.0018). Measurements on both devices were reproducible and of acceptable quality; the authors conclude home spirometry may not be interchanged with laboratory pneumotachography.

Watch out: Even professionally supervised, quality-acceptable home-device readings carried a systematic downward bias — supervision fixed technique, not device bias; children only, one named device.

Source: Brouwer AF, Roorda RJ, Brand PL. Comparison between peak expiratory flow and FEV(1) measurements on a home spirometer and on a pneumotachograph in children with asthma. Pediatric pulmonology. 2007. PMID: 17639585. Registry ID: not stated in the record. MeSH: Adolescent · Asthma · Child · Female · Forced Expiratory Volume · Humans · Male · Peak Expiratory Flow Rate · Reproducibility of Results · Spirometry

Card 4 — Synthesis — the setting, not the spirometer alone, decides agreement

Study: Synthesis

Population: Synthesis of the three studies carded on this shelf; no additional source.

Summary: Synthesis — the setting, not the spirometer alone, decides agreement

Result: Unsupervised home FEV1 in nearly 3,000 trial patients did not agree with clinic values well enough to interchange them, and a supervised home device in children still read systematically low against the laboratory standard. The single study that reached the individual-decision reliability standard (intraclass correlation 0.95-0.98) did so with virtual supervision plus human or AI over-reading in 68 patients over 8 days. Home spirometry is therefore a supervised-service measurement, not a drop-in replacement for the clinic number.

Watch out: Devices differ across cards (unnamed trial home spirometers, NuvoAir Air Next, Koko Peak Pro) and results are device- and protocol-specific; nothing here pools them.

Source: Synthesis of the carded records above (PMIDs 37385337, 41158491, 17639585). Registry ID: the source trials' ClinicalTrials.gov numbers are printed on the Oppenheimer card; not stated in the other records.