Automated 25-hydroxyvitamin D immunoassays versus LC-MS/MS (170 patient samples in the largest head-to-head, plus a 4-site assay evaluation and an LC-MS/MS method comparison): agreement is assay-specific, with biases from under 1 to over 10 nmol/L

Shelf #9739 · Vitamins & minerals

MEASUREMENT ACCURACY — NOT ESTABLISHED for interchangeable use across assays

Stamped: 2026-11-28.

Evidence published: 2010–2015.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

How accurately do the automated immunoassays used for 25-hydroxyvitamin D agree with liquid chromatography-tandem mass spectrometry (LC-MS/MS) — what bias and concordance does each platform show, and can results from different assays be treated as interchangeable for classifying vitamin D status?

Balance of the evidence

MEASUREMENT ACCURACY — NOT ESTABLISHED for interchangeable use across assays — Composition: 3 method-comparison studies (170 randomly selected patient samples in Farrell 2012; a 4-site evaluation of one competitive-binding assay against LC-MS/MS, HPLC, immunoassays and DEQAS pools in Wielders 2015, sample total not stated in its record; an LC-MS/MS versus radioimmunoassay and automated immunoassay comparison in van den Ouweland 2010, sample total not stated in its record). No sensitivity/specificity pair exists in these records — they report bias and concordance against the reference method, and none has been invented. The two LC-MS/MS methods agreed almost perfectly (concordance correlation 0.99, bias 0.56 µg/L, i.e. 1.4 nmol/L) and the radioimmunoassay was close (concordance 0.97, bias 1.1 µg/L, i.e. 2.7 nmol/L); among automated immunoassays, concordance ranged from 0.85 to 0.95 for total-25OH-D assays and mean bias from 0.2 µg/L (0.5 nmol/L) to 4.56 µg/L (11.4 nmol/L), while the D3-only Roche assay's concordance was just 0.66 (Farrell 2012). In the second comparison the DiaSorin radioimmunoassay agreed well with LC-MS/MS (r² = 0.90; average bias 1.61 nmol/L) while the Roche electrochemiluminescence assay disagreed considerably (r² = 0.58; bias 10.13 nmol/L) (van den Ouweland 2010). One modern binding assay showed LC-MS/MS-relative biases from −15.4% to +3.2% depending on the comparison method, and a −58.2% relative bias against one rival immunoassay (Wielders 2015). MEASUREMENT ACCURACY — NOT ESTABLISHED for interchangeable use: a 25-hydroxyvitamin D number belongs to its assay; the same serum can be classified differently by different platforms, and only LC-MS/MS-anchored methods agreed tightly in these head-to-heads.

MeSH: Automation · Chromatography, Liquid · Humans · Immunoassay · Sensitivity and Specificity · Tandem Mass Spectrometry · Vitamin D · Least-Squares Analysis · Luminescent Measurements · Radioimmunoassay · Reproducibility of Results · 25-Hydroxyvitamin D 2 · Binding, Competitive · Blood Chemical Analysis · Diagnostic Tests, Routine · Protein Binding · Vitamin D-Binding Protein

Cards (4)

Card 1 — Five automated immunoassays on 170 samples: concordance with LC-MS/MS from 0.66 to 0.95, bias up to 11.4 nmol/L (Farrell 2012)

Study: Primary method-comparison study (170 patient samples, 8 methods)

Population: 170 randomly selected patient samples, each measured by 2 LC-MS/MS methods, a DiaSorin radioimmunoassay, and automated immunoassays from Abbott, DiaSorin, IDS, Roche and Siemens; precision from 5 replicates of 2 serum pools on 5 days.

Summary: Five automated immunoassays on 170 samples: concordance with LC-MS/MS from 0.66 to 0.95, bias up to 11.4 nmol/L (Farrell 2012)

Result: The LC-MS/MS methods agreed with each other (concordance correlation coefficient 0.99; bias 0.56 µg/L, i.e. 1.4 nmol/L) and the radioimmunoassay performed comparably (concordance 0.97; bias 1.1 µg/L, i.e. 2.7 nmol/L). Among immunoassays measuring total 25-hydroxyvitamin D, concordance ranged from 0.85 (Abbott) to 0.95 (DiaSorin LIAISON) and mean bias from 0.2 µg/L (0.5 nmol/L, LIAISON) to 4.56 µg/L (11.4 nmol/L, Abbott). The Roche assay, which measures D3 only, had a small mean bias (−2.7 µg/L, i.e. −6.7 nmol/L) but a concordance of just 0.66. Most assays had coefficients of variation under 10%.

Watch out: Platform versions from 2012: manufacturers have since restandardised several assays, so the card ranks the problem (assay-specific bias) rather than today's exact league table. Its MeSH includes 'Sensitivity and Specificity' as a heading, but the study reports agreement statistics, not a diagnostic sensitivity/specificity pair.

Source: Farrell CJ, Martin S, McWhinney B et al.. State-of-the-art vitamin D assays: a comparison of automated immunoassays with liquid chromatography-tandem mass spectrometry methods. Clinical chemistry. 2012. PMID: 22230812. Registry ID: not stated in the record. MeSH: Automation · Chromatography, Liquid · Humans · Immunoassay · Sensitivity and Specificity · Tandem Mass Spectrometry · Vitamin D

Card 2 — Radioimmunoassay tracked LC-MS/MS (bias 1.61 nmol/L); the automated Roche assay did not (bias 10.13 nmol/L, r² = 0.58) (van den Ouweland 2010)

Study: Primary method-comparison study (sample total not stated in the record)

Population: Human serum samples measured by an isotope-dilution LC-MS/MS method for 25-hydroxyvitamin D2 and D3, the DiaSorin radioimmunoassay, and a restandardised Roche automated immunoassay; DEQAS external-quality samples also analysed (sample total not stated in the record).

Summary: Radioimmunoassay tracked LC-MS/MS (bias 1.61 nmol/L); the automated Roche assay did not (bias 10.13 nmol/L, r² = 0.58) (van den Ouweland 2010)

Result: The LC-MS/MS method was linear from 5 to at least 550 nmol/L with intra- and inter-day coefficients of variation of 6% or less, and its DEQAS pool results averaged −7.2% from the overall LC-MS method mean. The DiaSorin radioimmunoassay agreed well with LC-MS/MS (r² = 0.90; average bias 1.61 nmol/L); the Roche electrochemiluminescence assay disagreed considerably (r² = 0.58; bias 10.13 nmol/L).

Watch out: The Roche assay evaluated was a recently restandardised version of its era; the durable finding is the spread between platforms on identical samples, not a verdict on the current Roche assay.

Source: van den Ouweland JM, Beijers AM, Demacker PN, van Daal H. Measurement of 25-OH-vitamin D in human serum using liquid chromatography tandem-mass spectrometry with comparison to radioimmunoassay and automated immunoassay. Journal of chromatography. B, Analytical technologies in the biomedical and life sciences. 2010. PMID: 20381436. Registry ID: not stated in the record. MeSH: Automation · Chromatography, Liquid · Humans · Least-Squares Analysis · Luminescent Measurements · Radioimmunoassay · Reproducibility of Results · Sensitivity and Specificity · Tandem Mass Spectrometry · Vitamin D

Card 3 — One modern binding assay, four sites: −15.4% to +3.2% bias against LC-MS/MS methods and −58.2% against a rival immunoassay (Wielders 2015)

Study: Primary method-comparison / multicentre assay evaluation (4 sites; sample total not stated in the record)

Population: Four evaluation sites testing the Roche Elecsys Vitamin D Total competitive protein-binding assay (standardised against LC-MS/MS) for precision, inter-laboratory variability, functional sensitivity, and agreement with LC-MS/MS, HPLC, two immunoassays and 2 years of DEQAS returns (sample total not stated in the record).

Summary: One modern binding assay, four sites: −15.4% to +3.2% bias against LC-MS/MS methods and −58.2% against a rival immunoassay (Wielders 2015)

Result: Within-run coefficients of variation were 7% or less, within-laboratory under 9.5%, between-laboratory 10.1% or less, with functional sensitivity below 9.8 nmol/L. Agreement with HPLC showed a relative bias of −8.8%; against LC-MS/MS methods the relative biases were −15.4%, −13.5%, −10.2% and +3.2%; against immunoassays, +14.5% versus DiaSorin LIAISON and −58.2% versus IDS-iSYS. Two-year DEQAS mean was 102% of the all-laboratory trimmed mean, and bias on a certified reference panel was under 4%.

Watch out: A manufacturer-context evaluation of a single assay: its most useful datum for this shelf is the −58.2% figure — two automated assays in routine use can disagree by more than half on the same samples.

Source: Wielders JP, Carter GF, Eberl H et al.. Automated Competitive Protein-Binding Assay for Total 25-OH Vitamin D, Multicenter Evaluation and Practical Performance. Journal of clinical laboratory analysis. 2015. PMID: 25132191. Registry ID: not stated in the record. MeSH: 25-Hydroxyvitamin D 2 · Automation · Binding, Competitive · Blood Chemical Analysis · Chromatography, Liquid · Diagnostic Tests, Routine · Humans · Protein Binding · Tandem Mass Spectrometry · Vitamin D-Binding Protein

Card 4 — Synthesis — precise within an assay, not comparable across assays

Study: Synthesis

Population: Synthesis of the three studies carded on this shelf; no additional source.

Summary: Synthesis — precise within an assay, not comparable across assays

Result: Each platform is internally precise (coefficients of variation mostly under 10%), and LC-MS/MS methods agree with each other to about 1.4 nmol/L. The failure is between platforms: biases against LC-MS/MS run from about 1.6 nmol/L (radioimmunoassay) to 10–11 nmol/L (two automated immunoassays in these comparisons), concordance falls as low as 0.66, and two routine immunoassays differed by 58% relative on the same samples. A patient's category — deficient, insufficient, sufficient — can therefore change with the analyser, and serial results should stay on one method.

Watch out: No carded record reports classification accuracy (sensitivity/specificity in words) at a clinical cut-off such as 50 nmol/L; the shelf's NOT ESTABLISHED stamp rests on agreement and bias, which is the evidence that exists.

Source: Synthesis of the carded records above (PMIDs 22230812, 20381436, 25132191). Registry ID: not stated in the records.