Cordella pulmonary artery pressure sensor for heart-failure management — 456 patients in the single-arm PROACTIVE-HF trial and 15 in the first-in-human SIRONA study: the 6-month hospitalisation-or-death rate beat its performance goal (0.15 vs 0.43 events per patient), with no randomised comparison
Shelf #9749 · Heart, vessels, blood pressure & clotting
POSSIBLE (benefit)
Stamped: 2026-11-28.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In patients with chronic heart failure and class III symptoms, does management guided by the implanted Cordella pulmonary artery pressure sensor (seated pressures, Endotronix) reduce heart-failure hospitalisation and death, judged against performance goals in single-arm studies?
Balance of the evidence
POSSIBLE (benefit) — Composition: 1 large single-arm open-label trial (456 implanted patients) and 1 first-in-human feasibility study (15 patients); total documented N = 471. In PROACTIVE-HF the 6-month rate of heart-failure hospitalisation or all-cause mortality was 0.15 events per patient against a prespecified performance goal of 0.43 (p < 0.0001), with freedom from device- or system-related complications of 99.2% and from sensor failure of 99.8%. The trial was converted from a randomised to a single-arm design before completion, so no concurrent control group exists. Tier C: single-arm evidence only. POSSIBLE (benefit): the event rate is far below a historical performance goal and the sensor is accurate and safe in first-in-human testing, but the randomised evidence that exists for pressure-guided care belongs to a different sensor system (CardioMEMS) and cannot be transferred to this device. The tested product is the Cordella Pulmonary Artery Sensor System (Endotronix, Inc., Chicago, IL, USA), named as the publications give it.
MeSH: Aged · Female · Humans · Male · Middle Aged · Heart Failure · Hospitalization · Prospective Studies · Pulmonary Artery · Pulmonary Wedge Pressure · Stroke Volume · Aged, 80 and over · Blood Pressure Monitoring, Ambulatory · Cardiac Catheterization · Delivery of Health Care
Cards (3)
Card 1 — PROACTIVE-HF: 6-month heart-failure hospitalisation or death rate 0.15 events per patient versus a 0.43 performance goal; complications-free rate 99.2% (Guichard 2024, 456 patients)
Study: Single-arm trial
Population: 456 successfully implanted patients with chronic heart failure, NYHA class III symptoms, any ejection fraction, and a recent heart-failure hospitalisation and/or raised natriuretic peptides, at 75 centres in the USA and Europe; management used seated mean pulmonary artery pressure from the Cordella sensor (Endotronix) (NCT04089059).
Summary: PROACTIVE-HF: 6-month heart-failure hospitalisation or death rate 0.15 events per patient versus a 0.43 performance goal; complications-free rate 99.2% (Guichard 2024, 456 patients)
Result: The primary effectiveness endpoint at 6 months was met: the heart-failure hospitalisation or all-cause mortality rate was 0.15 events per patient (95% confidence interval 0.12 to 0.20), significantly below the 0.43 performance goal (p < 0.0001). Freedom from device- or system-related complications was 99.2% and freedom from pressure-sensor failure was 99.8% through 6 months.
Watch out: The trial was changed from randomised to single-arm after GUIDE-HF, with regulator input — the performance goal comes from earlier trials of a different sensor, in different eras of background therapy. Registry ID NCT04089059 is stated in the record.
Source: PMID: 39152983
Card 2 — SIRONA first-in-human: all 15 sensors implanted without failure or system complications, and sensor pressures tracked catheter pressures (cohort difference 2.7 mmHg at 90 days) (Mullens 2020, 15 patients)
Study: First-in-human feasibility study
Population: 15 patients with NYHA class III heart failure in a multicentre open-label first-in-human feasibility study of the Cordella Heart Failure System and Cordella Pulmonary Artery Pressure Sensor (Endotronix, Inc., Chicago, IL, USA — named as the publication gives it).
Summary: SIRONA first-in-human: all 15 sensors implanted without failure or system complications, and sensor pressures tracked catheter pressures (cohort difference 2.7 mmHg at 90 days) (Mullens 2020, 15 patients)
Result: All patients were successfully implanted without sensor failure, and no device system-related complications (invasive treatment, explant or death) occurred. The 90-day mean pulmonary artery pressure endpoint was met in all but one patient, with a cohort difference of 2.7 mmHg between sensor (22.5 ± 11.8 mmHg) and Swan-Ganz catheter (25.2 ± 8.5 mmHg). Adherence to daily measurement and transmission was 99%.
Watch out: Fifteen patients and an accuracy endpoint: this study establishes that the sensor measures pressure safely, not that acting on its readings changes outcomes. No registry ID is stated in the record.
Source: PMID: 32476191
Card 3 — Synthesis: accurate sensor, low event rate against a historical goal, no randomised evidence for this system
Study: Synthesis
Population: 471 patients in total across the two carded studies.
Summary: Synthesis: accurate sensor, low event rate against a historical goal, no randomised evidence for this system
Result: The Cordella programme shows a safe, accurate implantable sensor and a 6-month event rate well under a goal borrowed from earlier monitoring trials. Whether Cordella-guided management reduces hospitalisations compared with usual care — the question randomised trials answered, with mixed results, for CardioMEMS — is untested for this system in the carded records.
Watch out: Do not borrow the CardioMEMS randomised results (CHAMPION, GUIDE-HF, MONITOR-HF) as evidence for Cordella: different sensor, different (seated) measurement protocol, different management system.
Source: Synthesis of PMID: 39152983 and PMID: 32476191