SourceMed · Shelf #9778 in the ask interface

Bioabsorbable nasal implant (Latera) for nasal valve collapse (3 reports; 137 randomised): responders 82.5% versus 54.7% on sham at 3 months, with responder rates above 80% maintained uncontrolled to 24 months

Shelf #9778 · Eyes, skin, nose & allergy

POSSIBLE (benefit) — one industry sham-controlled randomised trial (137 patients) found a responder advantage of about 28 percentage points over a large sham response at 3 months, with uncontrolled durability to 24 months after crossover; the pooled literature is 5 studies and 396 patients, dominated by uncontrolled series

Stamped: 2026-11-28.

Evidence published: 2019–2022.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In adults with nasal obstruction from dynamic nasal valve collapse (lateral wall insufficiency), does a bioabsorbable lateral-wall implant improve nasal obstruction symptoms compared with a sham procedure?

Balance of the evidence

POSSIBLE (benefit) — Composition: 1 prospective multicentre single-blinded sham-controlled randomised trial (137 patients; 70 implant, 67 sham), its 24-month follow-up report (111 implanted participants including crossovers), and 1 systematic review/meta-analysis (5 studies, 396 patients; pools overlap the trial and are not added). At 3 months the responder rate (at least one severity-category improvement or at least a 20% relative reduction in Nasal Obstruction Symptom Evaluation score) was 82.5% on implant versus 54.7% on sham (p = 0.001); mean NOSE score fell -42.4 versus -22.7 (p < 0.0001) and visual analogue scale -39.0 versus -13.3 (p < 0.0001). Nineteen procedure/implant-related adverse events were reported in 17 patients, all resolving without sequelae. In the follow-up, NOSE responder rates stayed above 80% through 24 months with mean reductions of at least 30 points, but after the 3-month sham comparison the sham arm crossed over, so durability is uncontrolled; implant migration or retrieval affected 4.5% of implants (10 of 222). The meta-analysis likewise found improved quality-of-life scores versus sham and about a 5% incidence of mostly minor adverse effects. The tested product is the Latera bioabsorbable implant, named as the record names it for clarification; the record names no manufacturer and none is named here. POSSIBLE (benefit): a genuine short-term gain over a substantial sham response, from a single trial programme; whether the effect outlasts the implant's absorption is untested against any control.

MeSH: Absorbable Implants · Adult · Aged · Female · Humans · Male · Middle Aged · Nasal Obstruction · Nasal Surgical Procedures · Plastic Surgery Procedures · Single-Blind Method · Treatment Outcome · Follow-Up Studies · Nose · Randomized Controlled Trials as Topic

Cards (4)

Card 1 — Sham-controlled randomised trial, 137 patients: responders 82.5% on implant versus 54.7% on sham at 3 months (Stolovitzky 2019)

Study: Randomised sham-controlled trial

Population: 137 patients with dynamic nasal valve collapse at 10 clinics, randomised single-blind to the bioabsorbable implant (Latera, named in the record for clarification; no manufacturer is named in the record) placed in an in-office minimally invasive procedure (70 patients) or a sham procedure (67 patients), followed to 3 months.

Summary: Sham-controlled randomised trial, 137 patients: responders 82.5% on implant versus 54.7% on sham at 3 months (Stolovitzky 2019)

Result: Primary endpoint responder rate: 82.5% treatment versus 54.7% sham (p = 0.001). Mean NOSE score change -42.4 ± 23.4 versus -22.7 ± 27.9 (p < 0.0001); mean visual analogue scale change -39.0 ± 29.7 versus -13.3 ± 30.0 (p < 0.0001). Seventeen patients reported 19 procedure/implant-related adverse events, all resolved with no clinical sequelae.

Watch out: Single-blinded (patients), symptom-score outcomes, and a sham response above 50% mean the implant-specific gain is the roughly 28-point responder difference, not the 82.5% headline; the comparison ends at 3 months by design.

Source: Stolovitzky P et al. International Forum of Allergy & Rhinology. 2019. PMID: 31226238. Registry ID: not stated in the record. MeSH: Absorbable Implants · Adult · Aged · Female · Humans · Male · Middle Aged · Nasal Obstruction · Nasal Surgical Procedures · Plastic Surgery Procedures · Single-Blind Method · Treatment Outcome

Card 2 — 24-month follow-up of the same trial programme, 111 implanted participants: responder rates above 80% maintained, uncontrolled after crossover; retrieval in 4.5% of implants (Bikhazi 2022)

Study: Long-term follow-up report of the randomised trial (same programme — not a second trial)

Population: 111 participants with severe or extreme nasal airway obstruction primarily from nasal valve insufficiency who received implants in the trial's treatment and crossover arms, followed at 3, 6, 12, 18 and 24 months (90 completed 12 months; 70 completed 24 months).

Summary: 24-month follow-up of the same trial programme, 111 implanted participants: responder rates above 80% maintained, uncontrolled after crossover; retrieval in 4.5% of implants (Bikhazi 2022)

Result: NOSE responder rates were greater than 80% at every follow-up through 24 months, with mean NOSE reductions of at least 30 points (p < 0.001 at all time points) and mean visual analogue reductions of at least 29.7 points. No serious device- or procedure-related adverse events occurred; implant migration/retrieval was 4.5% of implants (10 of 222), 9% of participants (10 of 111).

Watch out: After crossover there is no control group, and 37% of implanted participants did not complete 24 months; sustained scores mix implant effect, natural history and completer selection.

Source: Bikhazi N et al. Facial Plastic Surgery. 2022. PMID: 34965603. Registry ID: not stated in the record. MeSH: Adult · Humans · Follow-Up Studies · Nasal Obstruction · Treatment Outcome · Nose · Plastic Surgery Procedures

Card 3 — Systematic review and meta-analysis, 5 studies / 396 patients: implant improved disease-specific quality of life versus sham; adverse effects about 5%, all resolved (Kim 2020)

Study: Systematic review and meta-analysis

Population: Systematic review of studies of the bioabsorbable nasal implant for nasal obstruction from lateral wall insufficiency, scoring endoscopic lateral wall movement and obstruction-related quality of life (5 studies, 396 patients; searches to September 2019).

Summary: Systematic review and meta-analysis, 5 studies / 396 patients: implant improved disease-specific quality of life versus sham; adverse effects about 5%, all resolved (Kim 2020)

Result: Implants significantly reduced endoscopic lateral wall motion versus pretreatment and improved quality of life at 12 months; against sham surgery, implants significantly improved disease-specific quality of life. Adverse effects (skin or mucosal reaction, infection, implant retrieval) were reported at about 5% incidence and all resolved without significant sequelae. The reviewers call for more randomised trials to verify effectiveness.

Watch out: Four of the 5 pooled studies are uncontrolled pre-post series, so the pooled symptom gains are not sham-adjusted; only the Stolovitzky trial inside the pool carries a control group.

Source: Kim DH et al. International Forum of Allergy & Rhinology. 2020. PMID: 32282129. Registry ID: not stated in the record. MeSH: Absorbable Implants · Humans · Nasal Obstruction · Nose · Randomized Controlled Trials as Topic · Plastic Surgery Procedures

Card 4 — Synthesis: a real 3-month advantage over sham in one trial programme; everything beyond 3 months is uncontrolled

Study: Synthesis

Population: Across the 3 reports carded above (137 randomised; 396 patients in the overlapping pooled literature).

Summary: Synthesis: a real 3-month advantage over sham in one trial programme; everything beyond 3 months is uncontrolled

Result: The sham-controlled difference (responders 82.5% versus 54.7%; NOSE -42.4 versus -22.7) supports a device effect on symptoms in dynamic nasal valve collapse. Durability claims rest on post-crossover follow-up with attrition, and no carded record compares the implant with functional rhinoplasty or other surgical correction.

Watch out: Population is dynamic collapse with severe/extreme obstruction; results do not extend to static stenosis or to obstruction mainly from septal deviation or turbinate hypertrophy (covered by other shelves).

Source: Synthesis of the reports carded above; no additional source.