Cardiotocography for predicting fetal acidaemia (150 traces read under 4 guidelines, plus a meta-analysis of computerised short-term variation): the best guideline's discrimination was an area under the curve of 0.66, short-term variation pooled at sensitivity 0.57 / specificity 0.81, and observers agreed only modestly (kappa 0.23–0.35)

Shelf #9839 · Women’s health

PREDICTIVE PERFORMANCE — NOT ESTABLISHED for stand-alone prediction of neonatal acidaemia

Stamped: 2026-11-28.

Evidence published: 2016–2022.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

How accurately does cardiotocography predict fetal or neonatal acidaemia — what sensitivity and specificity (labelled in words) do the international classification guidelines and computerised short-term variation achieve against cord-blood pH, and how well do observers agree when classifying the same traces?

Balance of the evidence

PREDICTIVE PERFORMANCE — NOT ESTABLISHED for stand-alone prediction of neonatal acidaemia — Composition: 1 diagnostic meta-analysis of computerised cardiotocography short-term variation (Kapaya 2016) and 1 four-guideline reader study of 150 traces (Zamora Del Pozo 2022). Sensitivity and specificity are labelled in words at first occurrence on this shelf: pooled short-term variation predicted fetal acidaemia with sensitivity 0.57 (95% CI 0.45 to 0.68) and specificity 0.81 (95% CI 0.69 to 0.89), positive likelihood ratio 3.14 and negative likelihood ratio 0.58, with a small improvement in growth-restricted fetuses (sensitivity 0.63) (Kapaya 2016). In the reader study, the last 30 minutes of 150 traces spanning all pH ranges were classified blindly by three reviewers under the FIGO, ACOG, NICE and Chandraharan guidelines: the Chandraharan classification had the highest sensitivity (78.79%) and the ACOG classification the highest specificity (95.73%), but the best discrimination was only moderate — area under the curve 0.66 (95% CI 0.55 to 0.77) — and did not differ significantly between guidelines; interobserver agreement (Fleiss kappa) was highest for FIGO at 0.35 (95% CI 0.28 to 0.41) and lowest for ACOG at 0.23 (95% CI 0.16 to 0.30) (Zamora Del Pozo 2022). PREDICTIVE PERFORMANCE — NOT ESTABLISHED for stand-alone prediction: every carded classification trades sensitivity against specificity at a modest overall discrimination, and the clinicians applying it agree only modestly on the same trace.

MeSH: Acidosis · Cardiotocography · Female · Gestational Age · Heart Rate, Fetal · Humans · Infant, Newborn · Pregnancy · Prenatal Diagnosis · ROC Curve · Reference Standards · Risk · Sensitivity and Specificity · Observer Variation

Cards (3)

Card 1 — Meta-analysis: computerised short-term variation predicts acidaemia at sensitivity 0.57 / specificity 0.81 (labelled in words) (Kapaya 2016)

Study: Systematic review and meta-analysis of diagnostic accuracy studies

Population: Systematic review and meta-analysis of studies of computerised cardiotocography short-term variation (STV) before delivery in singleton pregnancies at 24 weeks or more, against umbilical arterial pH in cord blood at birth; the record does not state the number of included studies or fetuses in the text carded here.

Summary: Meta-analysis: computerised short-term variation predicts acidaemia at sensitivity 0.57 / specificity 0.81 (labelled in words) (Kapaya 2016)

Result: Meta-analysis showed moderate accuracy of short-term variation in predicting fetal acidaemia: sensitivity 0.57 (95% CI 0.45 to 0.68) and specificity 0.81 (95% CI 0.69 to 0.89) — sensitivity and specificity labelled in words at first occurrence on this shelf — with a positive likelihood ratio of 3.14 (95% CI 2.13 to 4.63) and a negative likelihood ratio of 0.58 (95% CI 0.46 to 0.72). In intrauterine growth-restricted fetuses, detection improved slightly (sensitivity 0.63, 95% CI 0.49 to 0.75; negative likelihood ratio 0.50). The authors conclude short-term variation is a moderate predictor whose usefulness as a stand-alone test remains to be determined.

Watch out: This is the computerised antepartum/intrapartum short-term-variation measure, not a clinician's visual reading of the trace — the two are different index tests and the reader-study card scores the visual classifications separately.

Source: Kapaya H, Jacques R, Rahaim N et al.. "Does short-term variation in fetal heart rate predict fetal acidaemia?" A systematic review and meta-analysis. The journal of maternal-fetal & neonatal medicine : the official journal of the European Association of Perinatal Medicine, the Federation of Asia and Oceania Perinatal Societies, the International Society of Perinatal Obstetricians. 2016. PMID: 26902464. Registry ID: not stated in the record. MeSH: Acidosis · Cardiotocography · Female · Gestational Age · Heart Rate, Fetal · Humans · Infant, Newborn · Pregnancy · Prenatal Diagnosis · ROC Curve · Reference Standards · Risk · Sensitivity and Specificity

Card 2 — 150 traces, 4 guidelines, 3 blinded readers: best area under the curve 0.66, kappa only 0.23–0.35 (Zamora Del Pozo 2022)

Study: Reader / agreement study (150 traces, 3 reviewers)

Population: Reader study of the last 30 minutes of 150 cardiotocographic records spanning all pH ranges, each classified blindly by three independent reviewers under the FIGO, ACOG, NICE and Chandraharan guidelines, against neonatal acidaemia.

Summary: 150 traces, 4 guidelines, 3 blinded readers: best area under the curve 0.66, kappa only 0.23–0.35 (Zamora Del Pozo 2022)

Result: Sensitivity and specificity varied by guideline: the Chandraharan classification reached the highest sensitivity (78.79%) while ACOG had the highest specificity (95.73%). Averaged over the three observers, Chandraharan had the highest discrimination for neonatal acidaemia, but it was only moderate (area under the receiver operating characteristic curve 0.66, 95% CI 0.55 to 0.77) and did not differ significantly from the other guidelines. Interobserver agreement by Fleiss kappa was generally acceptable or moderate, highest with the FIGO classification (kappa 0.35, 95% CI 0.28 to 0.41) and lowest with ACOG (kappa 0.23, 95% CI 0.16 to 0.30).

Watch out: Kappa here measures whether readers classify the same trace alike, not whether the classification is right: a guideline can be consistently applied and still discriminate acidaemia only modestly, as the 0.66 area under the curve shows.

Source: Zamora Del Pozo C, Chóliz Ezquerro M, Mejía I et al.. Diagnostic capacity and interobserver variability in FIGO, ACOG, NICE and Chandraharan cardiotocographic guidelines to predict neonatal acidemia. The journal of maternal-fetal & neonatal medicine : the official journal of the European Association of Perinatal Medicine, the Federation of Asia and Oceania Perinatal Societies, the International Society of Perinatal Obstetricians. 2022. PMID: 34652249. Registry ID: not stated in the record. MeSH: Pregnancy · Infant, Newborn · Female · Humans · Heart Rate, Fetal · Observer Variation · Cardiotocography · Acidosis · Sensitivity and Specificity

Card 3 — Synthesis — modest discrimination, split by design: sensitive systems false-flag, specific systems miss

Study: Synthesis

Population: Synthesis of the two carded records; no additional source.

Summary: Synthesis — modest discrimination, split by design: sensitive systems false-flag, specific systems miss

Result: Both index tests land in the same place: computerised short-term variation pools at sensitivity 0.57 / specificity 0.81 (labelled in words), and the four visual guidelines trade the pair against each other (sensitivity up to 78.79%, specificity up to 95.73%) at a best area under the curve of 0.66, with readers agreeing at kappa 0.23–0.35 on identical traces. Cardiotocography stratifies risk; on the carded evidence it does not, on its own, establish or exclude fetal acidaemia.

Watch out: Boundary: whether continuous electronic fetal monitoring as a care strategy changes seizures, cerebral palsy or caesarean rates is live #4989 — an outcomes question; this shelf owns only how accurately the trace and its classifications predict acidaemia.

Source: Synthesis of the carded records above (PMIDs 26902464, 34652249). Registry ID: not stated in the records.