SourceMed · Shelf #9840 in the ask interface

Testosterone immunoassays versus mass spectrometry (238 people across the two comparison studies): biases of −90 and +99 ng/dL between platforms in men, mean results 46% above the reference in women's samples, and no immunoassay judged reliable at women's and children's concentrations

Shelf #9840 · Diabetes, glucose & metabolism

MEASUREMENT ACCURACY — NOT ESTABLISHED at low (women's, children's and hypogonadal) concentrations

Stamped: 2026-11-28.

Evidence published: 2003–2004.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

How accurately do routine testosterone immunoassays measure serum testosterone against liquid or gas chromatography–mass spectrometry — what bias and agreement do they show across the adult male range, and what happens at the low concentrations found in women, children and hypogonadal men?

Balance of the evidence

MEASUREMENT ACCURACY — NOT ESTABLISHED at low (women's, children's and hypogonadal) concentrations — Composition: 2 method-comparison studies (Wang 2004: 62 eugonadal and 60 hypogonadal men, 6 immunoassay methods against liquid chromatography–tandem mass spectrometry; Taieb 2003: 50 men, 55 women and 11 children, 10 immunoassays against isotope-dilution gas chromatography–mass spectrometry). No sensitivity/specificity pair arises on this shelf (a continuous assay compared by bias and agreement), and none is printed. In men, Deming regression slopes against mass spectrometry ran from 0.881 to 1.217 with correlation coefficients of 0.92 to 0.97, but platform biases were large in opposite directions (one automated platform averaging −90 ng/dL, another +99 ng/dL against the reference), and over 60% of samples in the adult male range fell within ±20% of the reference value (Wang 2004). At low concentrations the methods diverged: below 100 ng/dL (3.47 nmol/L) individual immunoassays overestimated, underestimated or scattered in both directions, and in women's samples 7 of 10 immunoassays overestimated testosterone, with mean results 46% above the mass-spectrometry values, while in men's samples they averaged 12% below; the authors judged none of the 10 immunoassays sufficiently reliable for sera from women and children (Taieb 2003). MEASUREMENT ACCURACY — NOT ESTABLISHED at low (women's, children's and hypogonadal) concentrations: immunoassays can separate eugonadal from hypogonadal men against locally established ranges, but at women's, children's and low hypogonadal concentrations their bias is platform-specific and large relative to the value being measured.

MeSH: Adult · Aged · Case-Control Studies · Chromatography, Liquid · Humans · Hypogonadism · Immunoassay · Male · Mass Spectrometry · Middle Aged · Regression Analysis · Reproducibility of Results · Testosterone · Autoanalysis · Child · Female · Gas Chromatography-Mass Spectrometry · Radioisotope Dilution Technique

Cards (3)

Card 1 — 122 men, 6 immunoassays versus LC-MS/MS: platform biases of −90 and +99 ng/dL despite correlations of 0.92–0.97 (Wang 2004)

Study: Method-comparison study (122 men, 7 methods)

Population: Method-comparison study in pedigreed serum samples from 62 eugonadal and 60 hypogonadal men, each measured by four automated immunoassay platforms and two manual immunoassay methods and by liquid chromatography–tandem mass spectrometry (LC-MS/MS).

Summary: 122 men, 6 immunoassays versus LC-MS/MS: platform biases of −90 and +99 ng/dL despite correlations of 0.92–0.97 (Wang 2004)

Result: Deming regression slopes for each method against LC-MS/MS ran from 0.881 to 1.217, and intraclass correlation coefficients were between 0.92 and 0.97 for all methods. Against the LC-MS/MS values, one automated platform (DPC Immulite) was biased toward lower values (mean difference −90 ± 9 ng/dL) and another (Bayer Centaur) toward higher values (+99 ± 11 ng/dL) across a wide range of concentrations. At low concentrations (below 100 ng/dL, printed in the record as 3.47 nmol/L), one manual assay overestimated, one automated platform underestimated, and two methods differed in both directions. Over 60% of samples within the adult male range were within ±20% of the LC-MS/MS result by most methods, and the authors conclude the immunoassays can distinguish eugonadal from hypogonadal males where each laboratory has established its own adult male reference ranges, but cannot accurately measure testosterone in females or prepubertal subjects.

Watch out: Platforms are named as published in 2004: the biases belong to those instrument-and-reagent generations and quantify what can go wrong between platforms, not the error of any current analyser.

Source: Wang C, Catlin DH, Demers LM et al.. Measurement of total serum testosterone in adult men: comparison of current laboratory methods versus liquid chromatography-tandem mass spectrometry. The Journal of clinical endocrinology and metabolism. 2004. PMID: 14764758. Registry ID: not stated in the record. MeSH: Adult · Aged · Case-Control Studies · Chromatography, Liquid · Humans · Hypogonadism · Immunoassay · Male · Mass Spectrometry · Middle Aged · Regression Analysis · Reproducibility of Results · Testosterone

Card 2 — 116 people, 10 immunoassays versus isotope-dilution GC-MS: means 46% too high in women, 12% too low in men (Taieb 2003)

Study: Method-comparison study (116 people, 11 methods)

Population: Method-comparison study measuring serum testosterone in 50 men, 55 women and 11 children with eight non-isotopic immunoassays, two isotopic immunoassays and isotope-dilution gas chromatography–mass spectrometry (ID/GC-MS) as reference.

Summary: 116 people, 10 immunoassays versus isotope-dilution GC-MS: means 46% too high in women, 12% too low in men (Taieb 2003)

Result: Compared with ID/GC-MS, 7 of the 10 immunoassays overestimated testosterone in samples from women, and mean immunoassay results were 46% above the reference values; in samples from men the immunoassays underestimated, with mean results 12% below the reference. In women (concentrations 0.6–7.2 nmol/L), 3 of the 10 immunoassays gave positive mean differences above 2.0 nmol/L (range −0.7 to 3.3 nmol/L); in men (8.2–58 nmol/L), 3 gave mean differences above 4.0 nmol/L (range −4.8 to 2.6 nmol/L). The authors conclude none of the immunoassays tested was sufficiently reliable for investigating sera from children and women, in whom very low (0.17 nmol/L) and low (below 1.7 nmol/L) concentrations are expected.

Watch out: A 46% mean overestimate in women is an average across assays and samples: individual assay-and-sample errors ran from −0.7 to +3.3 nmol/L, which at women's concentrations can exceed the value itself.

Source: Taieb J, Mathian B, Millot F et al.. Testosterone measured by 10 immunoassays and by isotope-dilution gas chromatography-mass spectrometry in sera from 116 men, women, and children. Clinical chemistry. 2003. PMID: 12881456. Registry ID: not stated in the record. MeSH: Adult · Aged · Autoanalysis · Child · Female · Gas Chromatography-Mass Spectrometry · Humans · Immunoassay · Male · Middle Aged · Radioisotope Dilution Technique · Testosterone

Card 3 — Synthesis — agreement is range-dependent: serviceable in the adult male range, unreliable where testosterone is low

Study: Synthesis

Population: Synthesis of the two carded records; no additional source.

Summary: Synthesis — agreement is range-dependent: serviceable in the adult male range, unreliable where testosterone is low

Result: Both comparisons, against two different mass-spectrometry references, agree on the pattern: correlations above 0.92 in men coexist with platform biases of about ±100 ng/dL, and accuracy collapses at low concentrations — immunoassay means 46% above the reference in women's samples, with errors in both directions below 100 ng/dL in hypogonadal men. A low testosterone result is therefore a property of the platform as well as the patient.

Watch out: Boundary: testosterone reference intervals from harmonised assays are pack #7397 (men) and pack #8473 (women), and daily testosterone production rates are pack #9587 — this shelf owns only how far routine immunoassay results sit from mass spectrometry.

Source: Synthesis of the carded records above (PMIDs 14764758, 12881456). Registry ID: not stated in the records.