Tonometers versus Goldmann applanation tonometry (documented N = 11,582 participants, 15,525 eyes, 102 studies): 95% limits of agreement from −3.8 to +4.3 mmHg for non-contact tonometry to −4.1 to +9.6 mmHg for the Ocuton S, with only 33–66% of readings within 2 mmHg of Goldmann

Shelf #9891 · Eyes, skin, nose & allergy

MEASUREMENT ACCURACY — PARTIAL

Stamped: 2026-11-28.

Evidence published: 2012.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

How closely do the tonometers used in clinical practice agree with Goldmann applanation tonometry, the reference device — what 95% limits of agreement and proportions within 2 mmHg does the systematic review find for each device type, and how large is observer variability?

Balance of the evidence

MEASUREMENT ACCURACY — PARTIAL — Composition: 1 systematic review and meta-analysis of directly comparative studies (102 studies, 130 paired comparisons, 11,582 participants, 15,525 eyes; Cook 2012) covering 8 tonometers against Goldmann applanation tonometry. No sensitivity/specificity pair arises on this shelf — intraocular pressure is a continuous measurement, so agreement (bias and 95% limits of agreement) is reported instead and no pair is invented. Summary 95% limits of agreement ran from a mean difference of 0.2 mmHg (limits −3.8 to 4.3 mmHg) for the non-contact tonometer to 2.7 mmHg (limits −4.1 to 9.6 mmHg) for the Ocuton S; the estimated proportion of readings within 2 mmHg of Goldmann ranged from 33% (Ocuton S) to 66% (non-contact) and 59% (handheld applanation). Substantial inter- and intraobserver variability were observed for all tonometers. MEASUREMENT ACCURACY — PARTIAL: the best devices centre close to Goldmann on average, but individual readings on every device can sit several mmHg away from it, so device readings are not interchangeable at the individual-patient level.

MeSH: Cross-Sectional Studies · Humans · Intraocular Pressure · Observer Variation · Predictive Value of Tests · Reproducibility of Results · Sensitivity and Specificity · Tonometry, Ocular

Cards (2)

Card 1 — 102 studies, 11,582 participants: agreement with Goldmann varies by device, widest for Ocuton S (Cook 2012)

Study: Systematic review and meta-analysis (102 studies, 11,582 participants)

Population: Systematic review and meta-analysis of directly comparative studies of 8 tonometers (dynamic contour, non-contact, ocular response analyzer, Ocuton S, handheld applanation, rebound, transpalpebral and Tono-Pen — named as published) against Goldmann applanation tonometry: 102 studies, 130 paired comparisons, 11,582 participants (15,525 eyes).

Summary: 102 studies, 11,582 participants: agreement with Goldmann varies by device, widest for Ocuton S (Cook 2012)

Result: Agreement (95% limits of agreement) varied across tonometers: from 0.2 mmHg (−3.8 to 4.3 mmHg) for the non-contact tonometer to 2.7 mmHg (−4.1 to 9.6 mmHg) for the Ocuton S. The estimated proportion within 2 mmHg of the Goldmann reading ranged from 33% (Ocuton S) to 66% and 59% (non-contact and handheld applanation tonometers respectively). The non-contact and handheld applanation tonometers achieved measurements closest to Goldmann, but there was substantial variability within and between studies, and substantial inter- and intraobserver variability for all tonometers.

Watch out: These are summary limits pooled across heterogeneous studies: a device's limits in any single clinic can be wider, and agreement with Goldmann is not the same as accuracy against true intraocular pressure — Goldmann is the accepted reference, not an error-free standard.

Source: Cook JA, Botello AP, Elders A et al.. Systematic review of the agreement of tonometers with Goldmann applanation tonometry. Ophthalmology. 2012. PMID: 22578443. Registry ID: not stated in the record. MeSH: Cross-Sectional Studies · Humans · Intraocular Pressure · Observer Variation · Predictive Value of Tests · Reproducibility of Results · Sensitivity and Specificity · Tonometry, Ocular

Card 2 — Synthesis — centred on Goldmann, scattered around it

Study: Synthesis

Population: Synthesis of the carded record; no additional source.

Summary: Synthesis — centred on Goldmann, scattered around it

Result: Across 102 comparative studies the average differences from Goldmann are small for the best devices (0.2 mmHg for non-contact tonometry) but the 95% limits of agreement span roughly ±4 mmHg even there and up to −4.1 to +9.6 mmHg for the worst-performing device, with as few as 33% of readings within 2 mmHg of the reference for that device. Device choice changes individual readings by clinically relevant amounts even when group means agree.

Watch out: Boundary: population intraocular pressure distributions are live #7454 and glaucoma treatment and surgery shelves are different questions — this shelf owns only device-versus-Goldmann agreement. No reliability threshold is invoked, so no 0.70/0.90 threshold explanation applies.

Source: Synthesis of the carded record above (PMID 22578443). Registry ID: not stated in the record.