Coronary artery calcium scoring, scan against rescan: mean interscan variability is about 12–22% overall but 61% at Agatston scores of 1–30 and only about 8–9% above 400 — the same score is a stable measurement at high burden and a noisy one at low burden
Shelf #9939 · Heart, vessels, blood pressure & clotting
MEASUREMENT ACCURACY — PARTIAL
Stamped: 2026-11-28.
Evidence published: 2002–2008.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In adults undergoing coronary artery calcium scanning, how reproducible is the score between two scans — what is the interscan variability of the Agatston and volume scores overall and by baseline score band, and which factors (heart-rate stability, ECG triggering, coronary motion, image noise) drive the variability?
Balance of the evidence
MEASUREMENT ACCURACY — PARTIAL — Composition: 2 paired duplicate-scan studies on electron-beam CT — 298 consecutive screening patients (Lu 2002) and 463 patients with end-stage renal disease in the CARE-2 study (Budoff 2008) — a documented base of 761 scanned patients. No sensitivity/specificity pair arises (a continuous score measured twice; reproducibility reported as published, none invented). Variability is strongly score-dependent: mean interscan variability was 21.6% (median 11.7%) for the Agatston score and 17.8% (median 10.8%) for the volume score in Lu 2002, falling from 34.6% at scores of 11–50 to 9.4% at 400–1,000; in CARE-2 the means were 11.8% (Agatston) and 10.3% (volume), but variability by band was 61.3% at scores 1–30, 23% at 31–100, 16.1% at 101–400 and 8.2% above 400 (Budoff 2008). Scan-to-scan correlation was 0.99 in both studies. MEASUREMENT ACCURACY — PARTIAL: highly reproducible at moderate-to-high scores, poorly reproducible in the low-score range where progression calls are most often made.
MeSH: Acetates · Adult · Aged · Aged, 80 and over · Atorvastatin · Calcinosis · Calcium Compounds · Chelating Agents · Coronary Artery Disease · Disease Progression · Drug Therapy, Combination · Electrocardiography · Female · Heart Rate · Heptanoic Acids · Humans · Hydroxymethylglutaryl-CoA Reductase Inhibitors · Kidney Failure, Chronic · Male · Middle Aged · Multivariate Analysis · Polyamines · Pyrroles · Reproducibility of Results · Sevelamer · Tomography, X-Ray Computed · Treatment Outcome
Cards (2)
Card 1 — Interscan variability 21.6% (Agatston) versus 17.8% (volume), shrinking as the score rises (Lu 2002)
Study: Paired duplicate-scan reproducibility study (N = 298)
Population: Two sets of electron-beam CT scans in 298 consecutive patients screened for coronary artery calcium, analysing interscan variation and its relation to heart rate, ECG triggering method, image noise and coronary motion.
Summary: Interscan variability 21.6% (Agatston) versus 17.8% (volume), shrinking as the score rises (Lu 2002)
Result: The interscan mean variabilities were 21.6% (median 11.7%) for the Agatston score and 17.8% (median 10.8%) for the volumetric score (P < .01); variability fell with rising score, from 34.6% at a score of 11–50 to 9.4% at 400–1,000 (P < .0001), and the absolute between-scan Agatston difference was 44.1 ± 95.6 with a scan-to-scan correlation of 0.99. Variability was lower in patients under 60, in those whose heart rate changed by less than 10 beats per minute, in those with no visible coronary motion, and with an optimal ECG triggering method.
Watch out: Electron-beam CT era technology: modern multidetector scanners may differ, but the score-dependence of the noise (worst at low scores) is a property of the score itself, not only of the scanner.
Source: Lu B, Budoff MJ, Zhuang N et al.. Causes of interscan variability of coronary artery calcium measurements at electron-beam CT. Academic radiology. 2002. PMID: 12061739. Registry ID: not stated in the record. MeSH: Adult · Aged · Aged, 80 and over · Calcinosis · Coronary Artery Disease · Electrocardiography · Female · Heart Rate · Humans · Male · Middle Aged · Multivariate Analysis · Reproducibility of Results · Tomography, X-Ray Computed
Card 2 — CARE-2 duplicate scans: 61.3% variability at scores 1–30, 8.2% above 400 (Budoff 2008)
Study: Paired duplicate-scan reproducibility study (N = 463)
Population: Duplicate electron-beam tomography scans at study entry in 463 patients with end-stage renal disease (mean age 59.4 ± 12.5 years, 48.3% female) in the CARE-2 study, with interscan variability assessed overall and across the clinical cut-point bands 1–30, 31–100, 101–400 and above 400.
Summary: CARE-2 duplicate scans: 61.3% variability at scores 1–30, 8.2% above 400 (Budoff 2008)
Result: Overall interscan variability was 11.8% for the Agatston score and 10.3% for the volume score. By band, Agatston variability was 61.3% (scores 1–30), 23% (31–100), 16.1% (101–400) and 8.2% (above 400); volume-score variability was 60.0%, 14.4%, 14.6% and 7.7% across the same bands. The scan-to-scan correlation rose with score, reaching 0.99 above 400 (P < .0001); the authors conclude reproducibility is sufficient for scores above 30.
Watch out: A dialysis population with heavy calcification: the band-specific pattern (worst at 1–30) matches the screening-population study, but absolute variability levels in general screening patients may differ.
Source: Budoff MJ, Kessler P, Gao YL et al.. The interscan variation of CT coronary artery calcification score: analysis of the Calcium Acetate Renagel Comparison (CARE)-2 study. Academic radiology. 2008. PMID: 18078907. Registry ID: not stated in the record. MeSH: Acetates · Atorvastatin · Calcinosis · Calcium Compounds · Chelating Agents · Coronary Artery Disease · Disease Progression · Drug Therapy, Combination · Female · Heptanoic Acids · Humans · Hydroxymethylglutaryl-CoA Reductase Inhibitors · Kidney Failure, Chronic · Male · Middle Aged · Polyamines · Pyrroles · Sevelamer · Tomography, X-Ray Computed · Treatment Outcome