SourceMed · Shelf #9988 in the ask interface
Automated 25-hydroxyvitamin D immunoassays against standardised LC-MS/MS: platforms disagree at the decision thresholds — one low-cost assay caught every deficient sample (sensitivity 100%) but over-called insufficiency, and the best-balanced platform still only reached kappa 0.82–0.90
Shelf #9988 · Vitamins & minerals
MEASUREMENT ACCURACY — PARTIAL
Stamped: 2026-11-28.
Evidence published: 2023–2026.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
How accurately do automated 25-hydroxyvitamin D immunoassays and newer automated mass-spectrometry analysers measure total 25-hydroxyvitamin D against certified / standardisation-traceable LC-MS/MS — what bias, limits of agreement and threshold classification (at 25, 50 and 75 nmol/L, i.e. 10, 20 and 30 ng/mL) do they show?
Balance of the evidence
MEASUREMENT ACCURACY — PARTIAL — Composition: 3 paired method-comparison studies against LC-MS/MS — 42 osteoporosis-clinic sera tested on a low-cost latex immunoturbidimetric assay, a chemiluminescent immunoassay and LC-MS/MS (Gören 2026); 400 outpatient sera across four concentration bands tested on three automated immunoassays against VDSCP-traceable LC-MS/MS (Racz 2026); and 124 sera tested on five methods including two mass-spectrometry systems against a certified LC-MS/MS (Favresse 2023). Sensitivity and specificity are labelled in words at first occurrence: the latex assay's sensitivity was 100.0% and specificity 86.7% for deficiency below 50 nmol/L (20 ng/mL), yet its slope against LC-MS/MS was 0.79 with mean bias −7.4 nmol/L (about −3.0 ng/mL) and limits of agreement −33.6 to +18.8 nmol/L, so it over-calls insufficiency (Gören 2026). Across platforms the failure mode differs by maker: the Roche platform overestimated (high specificity, reduced sensitivity), the Beckman platform caught the most deficiency but generated more false positives, and the DiaSorin platform balanced best with kappa 0.818–0.895 across thresholds (Racz 2026). An automated clinical LC-MS/MS analyser classified 95.2% of patients correctly and was not affected by the 3-epi metabolite that disturbed the immunoassays (Favresse 2023). MEASUREMENT ACCURACY — PARTIAL: results near a decision threshold are assay property, not patient property, and are not interchangeable between platforms.
MeSH: Adult · Automation · Chromatography, Liquid · Female · Humans · Immunoassay · Liquid Chromatography-Mass Spectrometry · Male · Middle Aged · Prospective Studies · Reproducibility of Results · Sensitivity and Specificity · Tandem Mass Spectrometry · Vitamin D · Vitamin D Deficiency
Cards (3)
Card 1 — Low-cost latex assay: sensitivity 100% / specificity 86.7% at 50 nmol/L, but slope 0.79 and bias −9.1% (Gören 2026)
Study: Paired method-comparison study (42 samples)
Population: 42 serum samples from osteoporosis-clinic outpatients analysed in parallel by a latex immunoturbidimetric assay (Beckman Coulter AU5800 — as named in the record), the Access 25(OH) Vitamin D Total chemiluminescent immunoassay (DxI 800) and LC-MS/MS as reference. The record carries no MeSH list.
Summary: Low-cost latex assay: sensitivity 100% / specificity 86.7% at 50 nmol/L, but slope 0.79 and bias −9.1% (Gören 2026)
Result: Sensitivity and specificity at the deficiency limit of 50 nmol/L (20 ng/mL) were sensitivity 100.0% / specificity 86.7% for the latex assay. Against LC-MS/MS its Passing–Bablok slope was 0.79 (95% CI 0.71 to 0.90), mean bias −7.4 nmol/L (−9.1%; about −3.0 ng/mL) with limits of agreement −33.6 to +18.8 nmol/L; 81.0% of its results fell within ±25% of LC-MS/MS (chemiluminescent assay 71.4%). Three-category agreement was 66.7% (kappa 0.51), every discordant result falling into a lower category — the assay over-calls insufficiency.
Watch out: Forty-two clinic samples: the perfect sensitivity is a small-sample finding, and the authors' own conclusion limits the assay to rule-out screening with LC-MS/MS confirmation near the limits.
Source: Gören Y, Şentüre Önder Ç et al.. Is a low-cost latex immunoturbidimetric assay suitable for routine 25(OH)D screening? Analytical comparison with chemiluminescent immunoassay and LC-MS/MS. Scandinavian journal of clinical and laboratory investigation. 2026. PMID: 42992596. Registry ID: not stated in the record. MeSH: not stated in the record (no MeSH list).
Card 2 — Three immunoassays, three failure modes at the same thresholds in 400 sera (Racz 2026)
Study: Paired method-comparison study (400 samples)
Population: 400 anonymised adult outpatient serum samples, equally distributed across four concentration bands (below 25, 25–49, 50–74, and at least 75 nmol/L — i.e. below 10, 10–19, 20–29, and at least 30 ng/mL), each measured on the DiaSorin Liaison XL, Beckman Coulter DxI 9000 and Roche cobas e602 immunoassays (as named in the record) and a VDSCP-traceable LC-MS/MS reference.
Summary: Three immunoassays, three failure modes at the same thresholds in 400 sera (Racz 2026)
Result: Significant analytical differences were found between every immunoassay and LC-MS/MS. The Roche platform consistently overestimated concentrations, giving reduced sensitivity but excellent specificity and positive predictive value across thresholds; the Beckman platform had the highest sensitivity for deficiency but more false-positive classifications, particularly at the 75 nmol/L (30 ng/mL) threshold; the DiaSorin platform achieved the best overall balance of sensitivity, specificity, accuracy and agreement, with Cohen's kappa 0.818 to 0.895 across decision thresholds. External quality-assessment (DEQAS) performance did not consistently reflect agreement with LC-MS/MS in patient samples.
Watch out: The record reports the platform ranking and kappa range without printing each platform's sensitivity/specificity pair — none is invented here; the carded finding is that the direction of error is a property of the platform.
Source: Racz S, Jozsa L, Bhattoa-Buzas AP et al.. Comprehensive Comparison of Three Automated 25-Hydroxyvitamin D Immunoassays with a VDSCP Traceable LC-MS/MS Method Across Clinically Relevant Decision Thresholds. Nutrients. 2026. PMID: 42796968. Registry ID: not stated in the record. MeSH: Humans · Vitamin D · Immunoassay · Tandem Mass Spectrometry · Liquid Chromatography-Mass Spectrometry · Vitamin D Deficiency · Reproducibility of Results · Chromatography, Liquid · Adult · Female · Automation · Middle Aged · Male · Sensitivity and Specificity · Prospective Studies
Card 3 — Automated LC-MS/MS analyser classified 95.2% correctly; the 3-epi metabolite disturbed only immunoassays (Favresse 2023)
Study: Paired method-comparison study (124 samples)
Population: 124 serum samples analysed by five methods — a reference LC-MS/MS, the automated Cascadion mass-spectrometry system, the Lumipulse assay and the Roche Elecsys II and III immunoassays (as named in the record) — with accuracy also checked against NIST SRM972a reference material and decision thresholds of 12, 20 and 30 ng/mL.
Summary: Automated LC-MS/MS analyser classified 95.2% correctly; the 3-epi metabolite disturbed only immunoassays (Favresse 2023)
Result: Against the reference LC-MS/MS, the Lumipulse and Roche Elecsys III assays reached the optimal bias criterion and the Cascadion the desirable criterion, while the Roche Elecsys II did not reach the minimal criterion. The proportion of correctly classified patients was higher with the Cascadion (95.2%) than with the three immunoassays, and the Cascadion was not affected by high 3-epi-25-hydroxyvitamin D3 concentrations, which disturbed the immunoassays, especially the Roche Elecsys methods.
Watch out: Correct classification is threshold-dependent and the immunoassay ranking here (Elecsys III acceptable) does not match every other study's platform order — another reason results are not portable between laboratories.
Source: Favresse J, Fangazio M, Cotton F et al.. Evaluation of four automated clinical analyzers for the determination of total 25(OH)D in comparison to a certified LC-MS/MS. Clinical chemistry and laboratory medicine. 2023. PMID: 36785905. Registry ID: not stated in the record. MeSH: Humans · Chromatography, Liquid · Tandem Mass Spectrometry · Vitamin D · Immunoassay