SourceMed · Shelf #9990 in the ask interface

Imaging tests for manifest glaucoma (OCT, scanning laser polarimetry, confocal scanning laser ophthalmoscopy): at specificity fixed near 0.95 the best parameters catch only about 0.67–0.72 of cases — average retinal nerve fibre layer sensitivity 0.69–0.78 across the reliable reviews

Shelf #9990 · Eyes, skin, nose & allergy

MEASUREMENT ACCURACY — PARTIAL

Stamped: 2026-11-28.

Evidence published: 2015–2021.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In people referred with suspected glaucoma (raised pressure, suspect disc or family history), how accurately do optical coherence tomography (OCT), scanning laser polarimetry (GDx) and confocal scanning laser ophthalmoscopy (HRT) detect manifest glaucoma against clinical diagnosis with visual fields — at specificity near 0.95, what sensitivity do the best parameters reach?

Balance of the evidence

MEASUREMENT ACCURACY — PARTIAL — Composition: a Cochrane diagnostic-accuracy review of 106 studies (16,260 eyes: 8,353 cases, 7,907 controls) (Michelessi 2015) and an overview of 4 systematic reviews of OCT accuracy (Michelessi 2021) — the overview's pools overlap the Cochrane primary literature, so the two are read as one evidence base read twice, not as independent confirmation. Sensitivity and specificity are labelled in words at first occurrence: at specificity held near 0.95, the best GDx parameter (Nerve Fibre Indicator) reached sensitivity 0.67 / specificity 0.94, the best HRT parameter (vertical cup/disc ratio) sensitivity 0.72 / specificity 0.94, and OCT average retinal nerve fibre layer thickness sensitivity 0.72 / specificity 0.93 (Michelessi 2015). The two reliable reviews in the overview put average RNFL sensitivity at 0.69 (0.63 to 0.73) and 0.78 (0.74 to 0.83) with specificity 0.94 and 0.93 (Michelessi 2021). At sensitivity 0.70 and specificity near 0.95, in 1,000 referred people of whom 200 have manifest glaucoma, the best measures would miss about 60 of the 200 cases and falsely refer about 50 of the 800 without glaucoma (Michelessi 2015). Accuracy was higher in more severe disease (sensitivity 0.79 at mean deviation worse than −6 dB vs 0.64 in milder disease), and 103 of the 106 studies were case-control designs, which can overestimate accuracy. Boundary: normal RNFL thickness values are pack #7179; this shelf is detection of manifest glaucoma only. MEASUREMENT ACCURACY — PARTIAL: imaging is specific but misses roughly three in ten manifest cases at that specificity, and the primary-study designs flatter it.

MeSH: Diagnostic Errors · Glaucoma · Humans · Intraocular Pressure · Nerve Fibers · Odds Ratio · Ophthalmoscopy · Optic Disk · Prospective Studies · Reproducibility of Results · Retrospective Studies · Scanning Laser Polarimetry · Sensitivity and Specificity · Tomography, Optical Coherence · Visual Field Tests · Visual Fields

Cards (2)

Card 1 — Cochrane: at specificity ~0.95, best imaging parameters reach sensitivity only 0.67–0.72 (Michelessi 2015)

Study: Cochrane diagnostic-accuracy review (106 studies, 16,260 eyes)

Population: 106 diagnostic-accuracy studies analysing 16,260 eyes (8,353 glaucoma cases, 7,907 controls) in people referred for suspected glaucoma: 40 studies of GDx scanning laser polarimetry (5,574 participants), 18 of HRT confocal scanning laser ophthalmoscopy (3,550 participants) and 63 of spectral-domain OCT (9,390 participants), with meta-analysis at specificity 0.95 or the closest reported level.

Summary: Cochrane: at specificity ~0.95, best imaging parameters reach sensitivity only 0.67–0.72 (Michelessi 2015)

Result: Sensitivity and specificity for the best parameter of each device: GDx Nerve Fibre Indicator sensitivity 0.67 (95% CI 0.55 to 0.77) / specificity 0.94 (0.92 to 0.95); HRT vertical cup/disc ratio sensitivity 0.72 / specificity 0.94; OCT average retinal nerve fibre layer thickness sensitivity 0.72 (0.65 to 0.77) / specificity 0.93 (0.92 to 0.95) — performance was remarkably similar across devices. Sensitivity was higher in more severe glaucoma (0.79 where mean deviation was worse than −6 dB vs 0.64 in milder disease). Of 106 studies, 103 used a case-control design, which can overestimate accuracy.

Watch out: Case-control spectra (clear cases vs clear controls) inflate accuracy relative to the consecutive referred patients in whom the test is actually used — real-world sensitivity at the same specificity will be lower.

Source: Michelessi M, Lucenteforte E, Oddone F et al.. Optic nerve head and fibre layer imaging for diagnosing glaucoma. The Cochrane database of systematic reviews. 2015. PMID: 26618332. Registry ID: not stated in the record. MeSH: Diagnostic Errors · Glaucoma · Humans · Nerve Fibers · Odds Ratio · Ophthalmoscopy · Optic Disk · Prospective Studies · Retrospective Studies · Scanning Laser Polarimetry · Sensitivity and Specificity · Tomography, Optical Coherence · Visual Field Tests

Card 2 — Overview of reviews: average RNFL sensitivity 0.69 and 0.78 at specificity 0.93–0.94 in the two reliable reviews (Michelessi 2021)

Study: Overview of systematic reviews (4 reviews; 50–57 studies in the reliable pair)

Population: Overview of 4 systematic reviews (published 2015–2018) of OCT for diagnosing glaucoma, each including between 17 and 113 primary studies; two reviews were at low risk of bias and two had high or unclear risk in several domains.

Summary: Overview of reviews: average RNFL sensitivity 0.69 and 0.78 at specificity 0.93–0.94 in the two reliable reviews (Michelessi 2021)

Result: The two reliable reviews reported accuracy of average retinal nerve fibre layer thickness as sensitivity 0.69 (0.63 to 0.73) with specificity 0.94 (0.93 to 0.95) in 57 studies, and sensitivity 0.78 (0.74 to 0.83) with specificity 0.93 (0.92 to 0.95) in 50 studies. Only one review clearly specified the clinical pathway, and both reliable reviews highlighted limitations of the primary diagnostic-accuracy studies.

Watch out: Single-synthesis caution flag in duplicate: this overview re-reads substantially the same primary literature as the Cochrane card, so agreement between the two cards is consistency of reading, not independent replication.

Source: Michelessi M, Li T, Miele A et al.. Accuracy of optical coherence tomography for diagnosing glaucoma: an overview of systematic reviews. The British journal of ophthalmology. 2021. PMID: 32493760. Registry ID: not stated in the record. MeSH: Glaucoma · Humans · Intraocular Pressure · Nerve Fibers · Optic Disk · Reproducibility of Results · Tomography, Optical Coherence · Visual Fields